Catalogue Search | MBRL
Search Results Heading
Explore the vast range of titles available.
MBRLSearchResults
-
DisciplineDiscipline
-
Is Peer ReviewedIs Peer Reviewed
-
Item TypeItem Type
-
SubjectSubject
-
YearFrom:-To:
-
More FiltersMore FiltersSourceLanguage
Done
Filters
Reset
1
result(s) for
"Multi-loop traction device"
Sort by:
Multi-loop traction device facilitates gastric endoscopic submucosal dissection: ex vivo pilot study and an inaugural clinical experience
by
Futakuchi, Toshiki
,
Kamba, Shunsuke
,
Sumiyama, Kazuki
in
Animals
,
Biopsy
,
Bowel and bladder training
2022
Background
Endoscopic submucosal dissection (ESD) is technically difficult and requires considerable training. The authors have developed a multi-loop traction device (MLTD), a new traction device that offers easy attachment and detachment. We aimed to evaluate the utility of MLTD in ESD.
Methods
This ex vivo pilot study was a prospective, block-randomized, comparative study of a porcine stomach model. Twenty-four lesions were assigned to a group that undertook ESD using the MLTD (M-ESD group) and a group that undertook conventional ESD (C-ESD group) to compare the speed of submucosal dissection. In addition, the data of consecutive 10 patients with eleven gastric lesions was collected using electronic medical records to clarify the inaugural clinical outcomes of gastric ESD using MLTD.
Results
The median (interquartile range) speed of submucosal dissection in the M-ESD and C-ESD groups were 141.5 (60.9–177.6) mm2/min and 35.5 (20.8–52.3) mm2/min, respectively; submucosal dissection was significantly faster in the M-ESD group (p < 0.05). The rate of en bloc resection and R0 resection was 100% in both groups, and there were no perforation in either group. The MLTD attachment time was 2.5 ± 0.9 min and the MLTD extraction time was 1.0 ± 1.1 min. Clinical outcomes of MLTD in gastric ESD were almost the same as those of ex vivo pilot study.
Conclusions
MLTD increased the speed of submucosal dissection in ESD and was similarly effective when used by expert and trainee endoscopists without perforation. MLTD can potentially ensure a safer and faster ESD.
Journal Article