Catalogue Search | MBRL
Search Results Heading
Explore the vast range of titles available.
MBRLSearchResults
-
DisciplineDiscipline
-
Is Peer ReviewedIs Peer Reviewed
-
Item TypeItem Type
-
SubjectSubject
-
YearFrom:-To:
-
More FiltersMore FiltersSourceLanguage
Done
Filters
Reset
140
result(s) for
"Radiation Dosimeters - standards"
Sort by:
Implementation of optically simulated luminescent dosimeter for quality control of gamma ray dose of an accelerator‐based neutron source
by
Keita Suga
,
Taiki Nakamura
,
Naonori Hu
in
accelerator neutron source
,
BNCT
,
Boron Neutron Capture Therapy - instrumentation
2024
Background Neutron beams utilized for performing BNCT are composed of a mixture of neutrons and gamma rays. Although much of the dose delivered to the cancer cells comes from the high LET particles produced by the boron neutron capture reaction, the dose delivered to the healthy tissues from unwanted gamma rays cannot be ignored. With the increase in the number of accelerators for BNCT, a detector system that is capable of measuring gamma ray dose in a mixed neutron/gamma irradiation field is crucial. Currently, BeO TLDs encased in quartz glass are used to measure gamma ray dose in a BNCT irradiation field. However, this type of TLD is no longer commercially available. A replacement dosimetry system is required to perform the recommended ongoing quality assurance of gamma ray measurement for a clinical BNCT system. Purpose The purpose of this study is to evaluate the characteristics of a BeO OSLD detector system under a mixed neutron and gamma ray irradiation field and to assess the suitability of the system for routine quality assurance measurements of an accelerator‐based BNCT facility. Methods The myOSLD system by RadPro International GmbH was evaluated using the accelerator‐based neutron source designed for clinical BNCT (NeuCure BNCT system). The readout constancy, linearity, dose rate effect, and fading effect of the OSLD were evaluated. Free‐in‐air and water phantom measurements were performed and compared with the TLD results and Monte Carlo simulation results. The PHITS Monte Carlo code was used for this study. Results The readout constancy was found to be stable over a month‐long period and similar to the TLD results. The OSLD readout signal was found to be linear, with a high coefficient of determination (R2 ≥ 0.999) up to a proton charge of 3.6 C. There was no significant signal fading or dose rate dependency. The central axis depth dose and off‐axis dose profile measurements agreed with both the TLD and Monte Carlo simulation results, within one standard deviation. Conclusion The myOSLD system was characterized using an accelerator system designed for clinical BNCT. The experimental measurements confirmed the OSLD achieved similar, if not superior to, the currently utilized dosimetry system for routine QA of an accelerator‐based BNCT system. The OSLD system would be a suitable replacement for the current TLD system for performing routine QA of gamma ray dose measurement in a BNCT irradiation field.
Journal Article
Assessing recall of personal sun exposure by integrating UV dosimeter and self-reported data with a network flow framework
by
Stump, Tammy K.
,
Alshurafa, Nabil
,
Spring, Bonnie
in
Activity recognition
,
Algorithms
,
Alignment
2019
Melanoma survivors often do not engage in adequate sun protection, leading to sunburn and increasing their risk of future melanomas. Melanoma survivors do not accurately recall the extent of sun exposure they have received, thus, they may be unaware of their personal UV exposure, and this lack of awareness may contribute towards failure to change behavior. As a means of determining behavioral accuracy of recall of sun exposure, this study compared subjective self-reports of time outdoors to an objective wearable sensor. Analysis of the meaningful discrepancies between the self-report and sensor measures of time outdoors was made possible by using a network flow algorithm to align sun exposure events recorded by both measures. Aligning the two measures provides the opportunity to more accurately evaluate false positive and false negative self-reports of behavior and understand participant tendencies to over- and under-report behavior.
39 melanoma survivors wore an ultraviolet light (UV) sensor on their chest while outdoors for 10 consecutive summer days and provided an end-of-day subjective self-report of their behavior while outdoors. A Network Flow Alignment framework was used to align self-report and objective UV sensor data to correct misalignment. The frequency and time of day of under- and over-reporting were identified.
For the 269 days assessed, the proposed framework showed a significant increase in the Jaccard coefficient (i.e. a measure of similarity between self-report and UV sensor data) by 63.64% (p < .001), and significant reduction in false negative minutes by 34.43% (p < .001). Following alignment of the measures, under-reporting of sun exposure time occurred on 51% of the days analyzed and more participants tended to under-report than to over-report sun exposure time. Rates of under-reporting of sun exposure were highest for events that began from 12-1pm, and second-highest from 5-6pm.
These discrepancies may reflect lack of accurate recall of sun exposure during times of peak sun intensity (10am-2pm) that could ultimately increase the risk of developing melanoma. This research provides technical contributions to the field of wearable computing, activity recognition, and identifies actionable times to improve participants' perception of their sun exposure.
Journal Article
Real-time quantitation of thyroidal radioiodine uptake in thyroid disease with monitoring by a collar detection device
by
Roussin, Jean-Paul
,
Solnes, Lilja
,
Nath, Tanmay
in
639/166/985
,
639/166/987
,
692/163/2743/1459
2021
Radioactive iodine (RAI) is safe and effective in most patients with hyperthyroidism but not all individuals are cured by the first dose, and most develop post-RAI hypothyroidism. Postoperative RAI therapy for remnant ablation is successful in 80–90% of thyroid cancer patients and sometimes induces remission of nonresectable cervical and/or distant metastatic disease but the effective tumor dose is usually not precisely known and must be moderated to avoid short- and long-term adverse effects on other tissues. The Collar Therapy Indicator (COTI) is a radiation detection device embedded in a cloth collar secured around the patient’s neck and connected to a recording and data transmission box. In previously published experience, the data can be collected at multiple time points, reflecting local cervical RAI exposure and correlating well with conventional methods. We evaluated the real-time uptake of RAI in patients with hyperthyroid Graves’ disease and thyroid cancer. We performed a pilot feasibility prospective study. Data were analyzed using R
©
(version 4.0.3, The R Foundation for Statistical Computing, 2020), and Python (version 3.6, Matplotlib version 3.0.3). The COTI was able to provide a quantitative temporal pattern of uptake within the thyroid in persons with Graves’ disease and lateralized the remnant tissue in persons with thyroid cancer. The study has demonstrated that the portable collar radiation detection device outside of a healthcare facility is accurate and feasible for use after administration of RAI for diagnostic studies and therapy to provide a complete collection of fractional target radioactivity data compared to that traditionally acquired with clinic-based measurements at one or two time-points.
Clinical Trials Registration
NCT03517579, DOR 5/7/2018.
Journal Article
First On-Site True Gamma-Ray Imaging-Spectroscopy of Contamination near Fukushima Plant
by
Tanimori, Toru
,
Tomono, Dai
,
Matsuoka, Yoshihiro
in
639/4077/4091/4094
,
639/766/930/2735
,
704/172/4081
2017
We have developed an Electron Tracking Compton Camera (ETCC), which provides a well-defined Point Spread Function (PSF) by reconstructing a direction of each gamma as a point and realizes simultaneous measurement of brightness and spectrum of MeV gamma-rays for the first time. Here, we present the results of our on-site pilot gamma-imaging-spectroscopy with ETCC at three contaminated locations in the vicinity of the Fukushima Daiichi Nuclear Power Plants in Japan in 2014. The obtained distribution of brightness (or emissivity) with remote-sensing observations is unambiguously converted into the dose distribution. We confirm that the dose distribution is consistent with the one taken by conventional mapping measurements with a dosimeter physically placed at each grid point. Furthermore, its imaging spectroscopy, boosted by Compton-edge-free spectra, reveals complex radioactive features in a quantitative manner around each individual target point in the background-dominated environment. Notably, we successfully identify a “micro hot spot” of residual caesium contamination even in an already decontaminated area. These results show that the ETCC performs exactly as the geometrical optics predicts, demonstrates its versatility in the field radiation measurement, and reveals potentials for application in many fields, including the nuclear industry, medical field, and astronomy.
Journal Article
A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis
by
Cheng, Kai-Yuan
,
Shih, Cheng-Ting
,
Hsieh, Ling-Ling
in
Acrylic Resins - chemistry
,
Biology and Life Sciences
,
Calibration
2016
Polymer gel dosimeters have been proven useful for dose evaluation in radiotherapy treatments. Previous studies have demonstrated that using a polymer gel dosimeter requires a 24 h reaction time to stabilize and further evaluate the measured dose distribution in two-dimensional dosimetry. In this study, the short-term stability within 24 h and feasibility of N-isopropylacrylamide (NIPAM) polymer gel dosimeters for use in three-dimensional dosimetry were evaluated using magnetic resonance imaging (MRI). NIPAM gels were used to measure the dose volume in a clinical case of intensity-modulated radiation therapy (IMRT). For dose readouts, MR images of irradiated NIPAM gel phantoms were acquired at 2, 5, 12, and 24 h after dose delivery. The mean standard errors of dose conversion from using dose calibration curves (DRC) were calculated. The measured dose volumes at the four time points were compared with those calculated using a treatment planning system (TPS). The mean standard errors of the dose conversion from using the DRCs were lower than 1 Gy. Mean pass rates of 2, 5, 12, and 24 h axial dose maps calculated using gamma evaluation with 3% dose difference and 3 mm distance-to-agreement criteria were 83.5% ± 0.9%, 85.9% ± 0.6%, 98.7% ± 0.3%, and 98.5% ± 0.9%, respectively. Compared with the dose volume histogram of the TPS, the absolute mean relative volume differences of the 2, 5, 12, and 24 h measured dose volumes were lower than 1% for the irradiated region with an absorbed dose higher than 2.8 Gy. It was concluded that a 12 h reaction time was sufficient to acquire accurate dose volume using the NIPAM gels with MR readouts.
Journal Article
Evaluation of basic characteristics of 3‐mm dose equivalent measuring instrument for evaluating lens exposure dose in radiotherapy
by
Isobe, Tomonori
,
Miyazaki, Shohei
,
Kobayashi, Daisuke
in
3‐mm dose equivalent
,
Accuracy
,
Dose-Response Relationship, Radiation
2023
Introduction Despite the development of DOSIRIS™, an eye lens dosimeter, the characteristics of DOSIRIS™ in the area of radiotherapy have not been investigated. The purpose of this study was to evaluate the basic characteristics of the 3‐mm dose equivalent measuring instrument DOSIRIS™ in radiotherapy. Methods Dose linearity and energy dependence were evaluated for the irradiation system based on the calibration method of the monitor dosimeter. The angle dependence was measured by irradiating from a total of 18 directions. Interdevice variation was repeated three times by simultaneously irradiating five dosimeters. The measurement accuracy was based on the absorbed dose measured by the monitor dosimeter of the radiotherapy equipment. Absorbed doses were converted to 3‐mm dose equivalents and compared with DOSIRIS™ measurements. Results Dose linearity was evaluated using the determination coefficient (R2) R2 = 0.9998 and 0.9996 at 6 and 10 MV, respectively. For energy dependence, although the therapeutic photons evaluated in this study had higher energies than in the previous studies and had a continuous spectrum, the response was equivalent to 0.2–1.25 MeV, well below the IEC 62387 limits. The maximum error at all angles was 15% (angle of 140°) and the coefficient of variation at all angles was 4.70%, which satisfies the standard of the thermoluminescent dosimeter measuring instrument. Accuracy of measurement was determined in terms of the measurement errors for DOSIRIS™ (3.2% and 4.3% at 6 and 10 MV, respectively,) using the 3‐mm dose equivalent obtained from the theoretical value as a reference. The DOSIRIS™ measurements met the IEC standard which defines the measurement error of ±30% of the irradiance value in IEC 62387. Conclusions We found that the characteristics of the 3‐mm dose equivalent dosimeter in a high‐energy radiation satisfy the IEC standards and have the same measurement accuracy as diagnostic areas such as Interventional Radiology. In 2012, the International Commission on Radiological Protection reported in their Publication 118 that the threshold dose of lens exposure has been significantly reduced and that the evaluation of lens exposure dose has emerged as a social issue. The purpose of the present study was to evaluate the basic characteristics of the 3‐mm dose equivalent measuring instrument DOSIRIS™ in radiotherapy.
Journal Article
Radiological tissue equivalence of deformable silicone‐based chemical radiation dosimeters (FlexyDos3D)
2019
FlexyDos3D, a silicone‐based chemical radiation dosimeter, has great potential to serve as a three‐dimensional (3D) deformable dosimetric tool to verify complex dose distributions delivered by modern radiotherapy techniques. To facilitate its clinical application, its radiological tissue needs to be clarified. In this study we investigated its tissue‐equivalence in comparison with water and Solid Water (RMI457). We found that its effective and mean atomic numbers were 40% and 20% higher and the total interaction probabilities for kV x‐ray photons were larger than those of water respectively. To assess the influence of its over‐response to kV photons, its HU value was measured by kV computed tomography (CT) and was found higher than all the soft‐tissue substitutes. When applied for dose calculation without correction, this effect led to an 8% overestimation in electron density via HU‐value mapping and 0.65% underestimation in target dose. Furthermore, depth dose curves (PDDs) and off‐axis ratios (profiles) at various beam conditions as well as the dose distribution of a full‐arc VMAT plan in FlexyDos3D and reference materials were simulated by Monte Carlo, where the results showed great agreement. As indicated, FlexyDos3D exhibits excellent radiological water‐equivalence for clinical MV x‐ray dosimetry, while its nonwater‐equivalent effect for low energy x‐ray dosimetry requires necessary correction. The key findings of this study provide pertinent reference for further FlexyDos3D characterization research.
Journal Article
Clinical radiation therapy measurements with a new commercial synthetic single crystal diamond detector
2014
A commercial version of a synthetic single crystal diamond detector (SCDD) in a Schottky diode configuration was recently released as the new type 60019 microDiamond detector (PTW‐Freiburg, Germany). In this study we investigate the dosimetric properties of this detector to independently confirm that findings from the developing group of the SCDDs still hold true for the commercial version of the SCDDs. We further explore if the use of the microDiamond detector can be expanded to high‐energy photon beams of up to 15 MV and to large field measurements. Measurements were performed with an Elekta Synergy linear accelerator delivering 6, 10, and 15 MV X‐rays, as well as 6, 9, 12, 15, and 20 MeV electron beams. The dependence of the microdiamond detector response on absorbed dose after connecting the detector was investigated. Furthermore, the dark current of the diamond detector was observed after irradiation. Results are compared to similar results from measurements with a diamond detector type 60003. Energy dependency was investigated, as well. Photon depth‐dose curves were measured for field sizes 3×3,10×10, and 30×30cm2. PDDs were measured with the Semiflex type 31010 detector, microLion type 31018 detector, P Diode type 60016, SRS Diode type 60018, and the microDiamond type 60019 detector (all PTW‐Freiburg). Photon profiles were measured at a depth of 10 cm. Electron depth‐dose curves normalized to the dose maximum were measured with the 14×14cm2 electron cone. PDDs were measured with a Markus chamber type 23343, an E Diode type 60017 and the microDiamond type 60019 detector (all PTW‐Freiburg). Profiles were measured with the E Diode and microDiamond at half of D90,D90,D70, and D50 depths and for electron cone sizes of 6×6cm2, 14×14cm2, and 20×20cm2. Within a tolerance of 0.5% detector response of the investigated detector was stable without any preirradiation. After preirradition with approximately 250 cGy the detector response was stable within 0.1%. A dark current after irradiation was not observed. The microDiamond detector shows no energy dependence in high energy photon or electron dosimetry. Electron PDD measurements with the E Diode and microDiamond are in good agreement. However, compared to E Diode measurements, dose values in the bremsstrahlungs region are about 0.5% lower when measured with the microDiamond detector. Markus detector measurements agree with E Diode measurements in the bremsstrahlungs region. For depths larger than dmax, depth‐dose curves of photon beams measured with the microDiamond detector are in close agreement to those measured with the microLion detector for small fields and with those measured with a Semiflex 0.125 cc ionization chamber for large fields. Differences are in the range of 0.25% and less. For profile measurements, microDiamond detector measurements agree well with microLion and P Diode measurements in the high‐dose region of the profile and the penumbra region. For areas outside the open field, P Diode measurements are about 0.5%–1.0% higher than microDiamond and microLion measurements. Thus it becomes evident that the investigated diamond detector (type 60019) is suitable for a wide range of applications in high‐energy photon and electron dosimetry and is interesting for relative, as well as absolute, dosimetry. PACS numbers: 00.06, 80.87
Journal Article
Water equivalent radiological properties of Gafchromic external beam therapy and external beam therapy 2 film dosimeters
2019
Background: Water equivalent property of any clinical dosimeter is important. Water has the approximately similar radiation absorption and scattering properties to soft tissue. Film dosimeter plays a significant role in radiotherapy quality assurance and treatment plan verification.
Aims: In this study, we are evaluating the water equivalent radiological properties of Gafchromic electronic benefit transfer (EBT) and EBT2 film dosimeters.
Materials and Methods: Radiological properties such as number of electrons per gram (ne), electron density (ϼe), and effective atomic number (Zeff) are calculated using Mayneord formula. Mixture rule is used to calculate the mass absorption coefficient (μen/ϼ) and mass attenuation coefficient (μ/ϼ), and data are generated using Win-XCom over 10 KeV to 20 MeV. Electron stopping power data are generated with the help of ESTAR database over 10 KeV to 30 MeV. Those results are compared with water and deviations are found.
Results: Our results suggest that Zeff, ne, and ϼe of EBT is showing deviations <8.83%, 4.39%, and 16.18% and for EBT2 is 4.26%, 2.82%, and 19.41% with respect to water. Deviation in μen/ϼ and μ/ϼ of EBT and EBT2 film is ≤5% and ≤6%, respectively, with respect to water >100 KeV. Electron stopping power properties are also close in agreements with water having deviations ≤5%.
Conclusion: Presence of high atomic number element chlorine, potassium, and bromine may disturb the water equivalent properties in the lower energy range <100 KeV and similarly enhance the dose sensitivity because of the strong photoelectric absorption process.
Journal Article
Evaluation of the effect of temperature variation on response of PRESAGE® dosimeter
by
Khezerloo, Najimeh
,
Hamrahi, Daryoosh
,
Nedaie, Hassan Ali
in
Biomedical engineering
,
Cancer
,
Cancer therapies
2017
Introduction:
Many factors, such as PRESAGE® composition, dose rate, energy, and type of radiation, temperature, etc., may effect on PRESAGE® dosimeter response. The aim of this study was investigating the effect of temperature variation on response of PRESAGE® solid dosimeter.
Materials and Methods:
In this study, a PRESAGE® solid detector was fabricated. Ninety-four percent weight polyurethane, 5% weight carbon tetrachloride, and 1% weight leucomalachite green were used. Radiological and physical characteristics of PRESAGEs®, such as mass density, electron density, and effective number atomic were obtained and compared with water. Response of PRESAGE® dosimeter in temperatures −4, 10, 25, 35, 45, 55, 65, 75, 85, and 90°C was evaluated. In addition, the absorption peak at various temperatures was investigated.
Results:
The results showed that the absorption peak at different temperatures was in the range of 630-635 nm. For temperatures below 75°C, the results indicated that temperature variation has no effect on the response of PRESAGE® dosimeter whereas at the temperatures >75°C, temperature variation has an effect on PRESAGE® dosimeter response.
Conclusion:
The finding showed that temperature changes have not impact on the absorption peak. In addition, the results related to the effect of temperature variation on the response of PRESAGE® dosimeter showed that in the range of clinical applications (temperatures below 75°C), temperature variation has no effect on PRESAGE® dosimeter response.
Journal Article