Patterns of adaptive servo-ventilation settings in a real-life multicenter study: pay attention to volume
by
Philippe, Carole
, Bourdin, Arnaud
, Georges, Marjolaine
, Palot, Alain
, Jaffuel, Dany
, Suehs, Carey M.
, Nogue, Erika
, Redolfi, Stefania
, Mallet, Jean-Pierre
, Rabec, Claudio
, Molinari, Nicolas
in
Medicine
/ Medicine & Public Health
/ Pneumology/Respiratory System
2020
Hey, we have placed the reservation for you!
By the way, why not check out events that you can attend while you pick your title.
You are currently in the queue to collect this book. You will be notified once it is your turn to collect the book.
Oops! Something went wrong.
Looks like we were not able to place the reservation. Kindly try again later.
Are you sure you want to remove the book from the shelf?
Patterns of adaptive servo-ventilation settings in a real-life multicenter study: pay attention to volume
by
Philippe, Carole
, Bourdin, Arnaud
, Georges, Marjolaine
, Palot, Alain
, Jaffuel, Dany
, Suehs, Carey M.
, Nogue, Erika
, Redolfi, Stefania
, Mallet, Jean-Pierre
, Rabec, Claudio
, Molinari, Nicolas
in
Medicine
/ Medicine & Public Health
/ Pneumology/Respiratory System
2020
Oops! Something went wrong.
While trying to remove the title from your shelf something went wrong :( Kindly try again later!
Do you wish to request the book?
Patterns of adaptive servo-ventilation settings in a real-life multicenter study: pay attention to volume
by
Philippe, Carole
, Bourdin, Arnaud
, Georges, Marjolaine
, Palot, Alain
, Jaffuel, Dany
, Suehs, Carey M.
, Nogue, Erika
, Redolfi, Stefania
, Mallet, Jean-Pierre
, Rabec, Claudio
, Molinari, Nicolas
in
Medicine
/ Medicine & Public Health
/ Pneumology/Respiratory System
2020
Please be aware that the book you have requested cannot be checked out. If you would like to checkout this book, you can reserve another copy
We have requested the book for you!
Your request is successful and it will be processed during the Library working hours. Please check the status of your request in My Requests.
Oops! Something went wrong.
Looks like we were not able to place your request. Kindly try again later.
Patterns of adaptive servo-ventilation settings in a real-life multicenter study: pay attention to volume
Journal Article
Patterns of adaptive servo-ventilation settings in a real-life multicenter study: pay attention to volume
2020
Request Book From Autostore
and Choose the Collection Method
Overview
Backgrounds
To explain the excess cardiovascular mortality observed in the SERVE-HF study, it was hypothesized that the high-pressure ASV default settings used lead to inappropriate ventilation, cascading negative consequences (i.e. not only pro-arrythmogenic effects through metabolic/electrolyte abnormalities, but also lower cardiac output). The aims of this study are: i) to describe ASV-settings for long-term ASV-populations in real-life conditions; ii) to describe the associated minute-ventilations (MV) and therapeutic pressures for servo-controlled-flow versus servo-controlled-volume devices (ASV-F Philips®-devices versus ASV-V ResMed®-devices).
Methods
The OTRLASV-study is a cross-sectional, 5-centre study including patients who underwent ASV-treatment for at least 1 year. The eight participating clinicians were free to adjust ASV settings, which were compared among i) initial diagnosed sleep-disordered-breathing (SBD) groups (Obstructive-Sleep-Apnea (OSA), Central-Sleep-Apnea (CSA), Treatment-Emergent-Central-Sleep-Apnea (TECSA)), and ii) unsupervised groups (
k
-means clusters). To generate these clusters, baseline and follow-up variables were used (age, sex, body mass index (BMI), initial diagnosed Obstructive-Apnea-Index, initial diagnosed Central-Apnea-Index, Continuous-Positive-Airway-Pressure used before ASV treatment, presence of cardiopathy, and presence of a reduced left-ventricular-ejection-fraction (LVEF)). ASV-data were collected using the manufacturer’s software for 6 months.
Results
One hundred seventy-seven patients (87.57% male) were analysed with a median (IQ
25–75
) initial Apnea-Hypopnea-Index of 50 (38–62)/h, an ASV-treatment duration of 2.88 (1.76–4.96) years, 61.58% treated with an ASV-V. SDB groups did not differ in ASV settings, MV or therapeutic pressures. In contrast, the five generated
k
-means clusters did (generally described as follows: (C1) male-TECSA-cardiopathy, (C2) male-mostly-CSA-cardiopathy, (C3) male-mostly-TECSA-no cardiopathy, (C4) female-mostly-elevated BMI-TECSA-cardiopathy, (C5) male-mostly-OSA-low-LVEF). Of note, the male-mostly-OSA-low-LVEF-cluster-5 had significantly lower fixed end-expiratory-airway-pressure (EPAP) settings versus C1 (
p
= 0.029) and C4 (
p
= 0.007). Auto-EPAP usage was higher in the male-mostly-TECSA-no cardiopathy-cluster-3 versus C1 (
p
= 0.006) and C2 (
p
< 0.001). MV differences between ASV-F (
p
= 0.002) and ASV-V (
p
< 0.001) were not homogenously distributed across clusters, suggesting specific cluster and ASV-algorithm interactions. Individual ASV-data suggest that the hyperventilation risk is not related to the cluster nor the ASV-monitoring type.
Conclusions
Real-life ASV settings are associated with combinations of baseline and follow-up variables wherein cardiological variables remain clinically meaningful. At the patient level, a hyperventilation risk exists regardless of cluster or ASV-monitoring type, spotlighting a future role of MV-telemonitoring in the interest of patient-safety.
Trial registration
The OTRLASV study was registered on ClinicalTrials.gov (Identifier:
NCT02429986
). 1 April 2015.