Compared Performance of High-Sensitivity Cameras Dedicated to Myocardial Perfusion SPECT: A Comprehensive Analysis of Phantom and Human Images
by
Franken, Philippe R.
, Verger, Antoine
, Songy, Bernard
, Poussier, Sylvain
, Karcher, Gilles
, Noel, Alain
, Morel, Olivier
, Wolf, Didier
, Marie, Pierre-Yves
, Imbert, Laetitia
in
Cameras
/ Cardiovascular disease
/ Gamma Cameras
/ Geometry
/ Humans
/ Male
/ Middle Aged
/ Myocardial Perfusion Imaging - instrumentation
/ Phantoms, Imaging
/ Sensitivity and Specificity
/ Tomography, Emission-Computed, Single-Photon - instrumentation
2012
Hey, we have placed the reservation for you!
By the way, why not check out events that you can attend while you pick your title.
You are currently in the queue to collect this book. You will be notified once it is your turn to collect the book.
Oops! Something went wrong.
Looks like we were not able to place the reservation. Kindly try again later.
Are you sure you want to remove the book from the shelf?
Compared Performance of High-Sensitivity Cameras Dedicated to Myocardial Perfusion SPECT: A Comprehensive Analysis of Phantom and Human Images
by
Franken, Philippe R.
, Verger, Antoine
, Songy, Bernard
, Poussier, Sylvain
, Karcher, Gilles
, Noel, Alain
, Morel, Olivier
, Wolf, Didier
, Marie, Pierre-Yves
, Imbert, Laetitia
in
Cameras
/ Cardiovascular disease
/ Gamma Cameras
/ Geometry
/ Humans
/ Male
/ Middle Aged
/ Myocardial Perfusion Imaging - instrumentation
/ Phantoms, Imaging
/ Sensitivity and Specificity
/ Tomography, Emission-Computed, Single-Photon - instrumentation
2012
Oops! Something went wrong.
While trying to remove the title from your shelf something went wrong :( Kindly try again later!
Do you wish to request the book?
Compared Performance of High-Sensitivity Cameras Dedicated to Myocardial Perfusion SPECT: A Comprehensive Analysis of Phantom and Human Images
by
Franken, Philippe R.
, Verger, Antoine
, Songy, Bernard
, Poussier, Sylvain
, Karcher, Gilles
, Noel, Alain
, Morel, Olivier
, Wolf, Didier
, Marie, Pierre-Yves
, Imbert, Laetitia
in
Cameras
/ Cardiovascular disease
/ Gamma Cameras
/ Geometry
/ Humans
/ Male
/ Middle Aged
/ Myocardial Perfusion Imaging - instrumentation
/ Phantoms, Imaging
/ Sensitivity and Specificity
/ Tomography, Emission-Computed, Single-Photon - instrumentation
2012
Please be aware that the book you have requested cannot be checked out. If you would like to checkout this book, you can reserve another copy
We have requested the book for you!
Your request is successful and it will be processed during the Library working hours. Please check the status of your request in My Requests.
Oops! Something went wrong.
Looks like we were not able to place your request. Kindly try again later.
Compared Performance of High-Sensitivity Cameras Dedicated to Myocardial Perfusion SPECT: A Comprehensive Analysis of Phantom and Human Images
Journal Article
Compared Performance of High-Sensitivity Cameras Dedicated to Myocardial Perfusion SPECT: A Comprehensive Analysis of Phantom and Human Images
2012
Request Book From Autostore
and Choose the Collection Method
Overview
Differences in the performance of cadmium-zinc-telluride (CZT) cameras or collimation systems that have recently been commercialized for myocardial SPECT remain unclear. In the present study, the performance of 3 of these systems was compared by a comprehensive analysis of phantom and human SPECT images.
We evaluated the Discovery NM 530c and DSPECT CZT cameras, as well as the Symbia Anger camera equipped with an astigmatic (IQ x SPECT) or parallel-hole (conventional SPECT) collimator. Physical performance was compared on reconstructed SPECT images from a phantom and from comparable groups of healthy subjects.
Classifications were as follows, in order of performance. For count sensitivity on cardiac phantom images (counts x s(-1) x MBq(-1)), DSPECT had a sensitivity of 850; Discovery NM 530c, 460; IQ x SPECT, 390; and conventional SPECT, 130. This classification was similar to that of myocardial counts normalized to injected activities from human images (respective mean values, in counts x s(-1) x MBq(-1): 11.4 ± 2.6, 5.6 ± 1.4, 2.7 ± 0.7, and 0.6 ± 0.1). For central spatial resolution: Discovery NM 530c was 6.7 mm; DSPECT, 8.6 mm; IQ x SPECT, 15.0 mm; and conventional SPECT, 15.3 mm, also in accordance with the analysis of the sharpness of myocardial contours on human images (in cm(-1): 1.02 ± 0.17, 0.92 ± 0.11, 0.64 ± 0.12, and 0.65 ± 0.06, respectively). For contrast-to-noise ratio on the phantom: Discovery NM 530c had a ratio of 4.6; DSPECT, 4.1; IQ x SPECT, 3.9; and conventional SPECT, 3.5, similar to ratios documented on human images (5.2 ± 1.0, 4.5 ± 0.5, 3.9 ± 0.6, and 3.4 ± 0.3, respectively).
The performance of CZT cameras is dramatically higher than that of Anger cameras, even for human SPECT images. However, CZT cameras differ in that spatial resolution and contrast-to-noise ratio are better with the Discovery NM 530c, whereas count sensitivity is markedly higher with the DSPECT.