Analytical and Clinical Considerations in Implementing the Roche Elecsys Troponin T Gen 5 STAT Assay
by
Baumann, Nikola A
, Jaffe, Allan S
, Donato, Leslie J
, Wockenfus, Amy M
, Karon, Brad S
, Katzman, Brooke M
in
Biotin
/ Chemical tests and reagents
/ Comparative analysis
/ Diagnostic Tests, Routine
/ Electronic components industry
/ Emergency medical care
/ Emergency Service, Hospital
/ Hemolysis
/ Hospital patients
/ Humans
/ Patients
/ Physical instruments
/ Reagents
/ Reference Values
/ Troponin
/ Troponin T
2021
Hey, we have placed the reservation for you!
By the way, why not check out events that you can attend while you pick your title.
You are currently in the queue to collect this book. You will be notified once it is your turn to collect the book.
Oops! Something went wrong.
Looks like we were not able to place the reservation. Kindly try again later.
Are you sure you want to remove the book from the shelf?
Analytical and Clinical Considerations in Implementing the Roche Elecsys Troponin T Gen 5 STAT Assay
by
Baumann, Nikola A
, Jaffe, Allan S
, Donato, Leslie J
, Wockenfus, Amy M
, Karon, Brad S
, Katzman, Brooke M
in
Biotin
/ Chemical tests and reagents
/ Comparative analysis
/ Diagnostic Tests, Routine
/ Electronic components industry
/ Emergency medical care
/ Emergency Service, Hospital
/ Hemolysis
/ Hospital patients
/ Humans
/ Patients
/ Physical instruments
/ Reagents
/ Reference Values
/ Troponin
/ Troponin T
2021
Oops! Something went wrong.
While trying to remove the title from your shelf something went wrong :( Kindly try again later!
Do you wish to request the book?
Analytical and Clinical Considerations in Implementing the Roche Elecsys Troponin T Gen 5 STAT Assay
by
Baumann, Nikola A
, Jaffe, Allan S
, Donato, Leslie J
, Wockenfus, Amy M
, Karon, Brad S
, Katzman, Brooke M
in
Biotin
/ Chemical tests and reagents
/ Comparative analysis
/ Diagnostic Tests, Routine
/ Electronic components industry
/ Emergency medical care
/ Emergency Service, Hospital
/ Hemolysis
/ Hospital patients
/ Humans
/ Patients
/ Physical instruments
/ Reagents
/ Reference Values
/ Troponin
/ Troponin T
2021
Please be aware that the book you have requested cannot be checked out. If you would like to checkout this book, you can reserve another copy
We have requested the book for you!
Your request is successful and it will be processed during the Library working hours. Please check the status of your request in My Requests.
Oops! Something went wrong.
Looks like we were not able to place your request. Kindly try again later.
Analytical and Clinical Considerations in Implementing the Roche Elecsys Troponin T Gen 5 STAT Assay
Journal Article
Analytical and Clinical Considerations in Implementing the Roche Elecsys Troponin T Gen 5 STAT Assay
2021
Request Book From Autostore
and Choose the Collection Method
Overview
Abstract
Objectives
To evaluate the analytical and clinical performance characteristics of the fifth-generation troponin T reagent.
Methods
Troponin T was measured in 2,332 paired serum and plasma samples from emergency department and hospital patients using the fourth- and fifth-generation reagents. Testing was repeated after recentrifugation to determine the frequency of analytical outliers and percentage of patients with elevated values for each assay. We conducted separate experiments to determine the effects of biotin and hemolysis interference, as well as measure interinstrument variability, for fifth-generation troponin T.
Results
Analytic outliers occurred more frequently using the fifth-generation reagent (3.4%) compared with the fourth-generation reagent (1.0%). The frequency of elevated troponin T above the 99th percentile upper reference limit was 26% for the fourth-generation reagent and 52% for the fifth-generation reagent. Clinically significant assay interference by biotin was observed at 20 ng/mL, but hemolysis interference was not observed until an H index of 150. Instrument-to-instrument variability between e411 and e601/602 instrument platforms is predicted to confound clinical interpretation of troponin changes.
Conclusions
Analytical outliers and instrument-to-instrument variability are the two analytical variables most likely to confound interpretation of changes in fifth-generation troponin T results over time.