Hip stability parameters with dual mobility, modular dual mobility and fixed bearing in total hip arthroplasty: an analytical evaluation
by
Valtorta, Riccardo
, Amendola, Luca
, Tigani, Domenico
, Banci, Lorenzo
in
Arthroplasty, Replacement, Hip
/ Artificial hip joints
/ Dislocation
/ Dual mobility
/ Epidemiology
/ Hip
/ Hip Dislocation - surgery
/ Hip Prosthesis
/ Humans
/ Internal Medicine
/ Joint replacement surgery
/ Jumping distance
/ Mechanical properties
/ Medicine
/ Medicine & Public Health
/ Mobility
/ Modular dual mobility
/ Musculoskeletal diseases
/ Orthopedics
/ Oscillation angle
/ Polyethylene
/ Prostheses
/ Prosthesis Design
/ Prosthesis Failure
/ Rehabilitation
/ Reoperation
/ Rheumatology
/ Sports Medicine
/ Titanium alloys
/ Total hip arthroplasty
/ Transplants & implants
2022
Hey, we have placed the reservation for you!
By the way, why not check out events that you can attend while you pick your title.
You are currently in the queue to collect this book. You will be notified once it is your turn to collect the book.
Oops! Something went wrong.
Looks like we were not able to place the reservation. Kindly try again later.
Are you sure you want to remove the book from the shelf?
Hip stability parameters with dual mobility, modular dual mobility and fixed bearing in total hip arthroplasty: an analytical evaluation
by
Valtorta, Riccardo
, Amendola, Luca
, Tigani, Domenico
, Banci, Lorenzo
in
Arthroplasty, Replacement, Hip
/ Artificial hip joints
/ Dislocation
/ Dual mobility
/ Epidemiology
/ Hip
/ Hip Dislocation - surgery
/ Hip Prosthesis
/ Humans
/ Internal Medicine
/ Joint replacement surgery
/ Jumping distance
/ Mechanical properties
/ Medicine
/ Medicine & Public Health
/ Mobility
/ Modular dual mobility
/ Musculoskeletal diseases
/ Orthopedics
/ Oscillation angle
/ Polyethylene
/ Prostheses
/ Prosthesis Design
/ Prosthesis Failure
/ Rehabilitation
/ Reoperation
/ Rheumatology
/ Sports Medicine
/ Titanium alloys
/ Total hip arthroplasty
/ Transplants & implants
2022
Oops! Something went wrong.
While trying to remove the title from your shelf something went wrong :( Kindly try again later!
Do you wish to request the book?
Hip stability parameters with dual mobility, modular dual mobility and fixed bearing in total hip arthroplasty: an analytical evaluation
by
Valtorta, Riccardo
, Amendola, Luca
, Tigani, Domenico
, Banci, Lorenzo
in
Arthroplasty, Replacement, Hip
/ Artificial hip joints
/ Dislocation
/ Dual mobility
/ Epidemiology
/ Hip
/ Hip Dislocation - surgery
/ Hip Prosthesis
/ Humans
/ Internal Medicine
/ Joint replacement surgery
/ Jumping distance
/ Mechanical properties
/ Medicine
/ Medicine & Public Health
/ Mobility
/ Modular dual mobility
/ Musculoskeletal diseases
/ Orthopedics
/ Oscillation angle
/ Polyethylene
/ Prostheses
/ Prosthesis Design
/ Prosthesis Failure
/ Rehabilitation
/ Reoperation
/ Rheumatology
/ Sports Medicine
/ Titanium alloys
/ Total hip arthroplasty
/ Transplants & implants
2022
Please be aware that the book you have requested cannot be checked out. If you would like to checkout this book, you can reserve another copy
We have requested the book for you!
Your request is successful and it will be processed during the Library working hours. Please check the status of your request in My Requests.
Oops! Something went wrong.
Looks like we were not able to place your request. Kindly try again later.
Hip stability parameters with dual mobility, modular dual mobility and fixed bearing in total hip arthroplasty: an analytical evaluation
Journal Article
Hip stability parameters with dual mobility, modular dual mobility and fixed bearing in total hip arthroplasty: an analytical evaluation
2022
Request Book From Autostore
and Choose the Collection Method
Overview
Background
Use of dual mobility (DM) in total hip arthroplasty has gained popularity due to the ability to reduce dislocation through increased jumping distance and impingement-free arc of movement. Recently, modular dual mobility (modDM) systems were introduced to give the possibility to use DM with standard metal-backed shells, however few has been studied to date regarding how jumping distance and the center of rotation change with modDM. The objective of this study was to evaluate, through analytical simulation, how jumping distance, center of rotation and arc of movement change between DM and standard cups with modDM or fixed bearings (FB).
Methods
3D-models of DM and standard press-fit cups with modDM or FB liners were used to simulate DM, modDM and FB implant configurations, matched for same cup size, according to same cup position and different femoral head diameters. Jumping distance was calculated and center of rotation lateralization and oscillation angles were measured for each size of these three implant configurations.
Results
Jumping distance with modDM was reduced by -3.9 mm to -8.6 mm in comparison with DM, from 48 to 64 mm size, but resulted comparable to polyethylene 36 mm FB and increased by + 1.1 mm and + 1.4 mm than ceramic 36 and 40 mm FBs for sizes > 54 mm. ModDM lateralized the center of rotation up to + 2.5 mm and + 4.0 mm in comparison with DM and FBs, respectively. Oscillation angle with modDM resulted higher than + 16°, + 23°, + 17° and + 14° in comparison to DM, 28 mm, 32 mm and 36 mm FB cups, respectively, for 56 mm cup size.
Conclusions
According to its specific design, modDM might change hip stability parameters in comparison to DM, worsening jumping distance and center of rotation position, but increasing arc of movement. As not restoring stability parameters in the same fashion, modDM implants should be properly used when DM cups are not feasible.