Asset Details
MbrlCatalogueTitleDetail
Do you wish to reserve the book?
A critical phantom study of the energy window used for 99m Tc quantitative explorations with a ring CZT SPECT system
by
Bernard, Claire
, Seret, Alain
2025
Hey, we have placed the reservation for you!
By the way, why not check out events that you can attend while you pick your title.
You are currently in the queue to collect this book. You will be notified once it is your turn to collect the book.
Oops! Something went wrong.
Looks like we were not able to place the reservation. Kindly try again later.
Are you sure you want to remove the book from the shelf?
Oops! Something went wrong.
While trying to remove the title from your shelf something went wrong :( Kindly try again later!
Do you wish to request the book?
A critical phantom study of the energy window used for 99m Tc quantitative explorations with a ring CZT SPECT system
by
Bernard, Claire
, Seret, Alain
2025
Please be aware that the book you have requested cannot be checked out. If you would like to checkout this book, you can reserve another copy
We have requested the book for you!
Your request is successful and it will be processed during the Library working hours. Please check the status of your request in My Requests.
Oops! Something went wrong.
Looks like we were not able to place your request. Kindly try again later.
A critical phantom study of the energy window used for 99m Tc quantitative explorations with a ring CZT SPECT system
Journal Article
A critical phantom study of the energy window used for 99m Tc quantitative explorations with a ring CZT SPECT system
2025
Request Book From Autostore
and Choose the Collection Method
Overview
This study investigated, mainly for quantitative
Tc explorations with a ring CZT SPECT system (GE HealthCare Starguide), the use of a narrow symmetric or a fully asymmetric energy window to reject scattered photons. The results were compared with the manufacturer's post-acquisition dual energy window approach.
Two uniform and two cold and hot rod contrast cylindrical phantoms of various sizes were scanned with the Starguide system to acquire a very high number of counts. After rebinning the list-mode files for different energy windows, data were reconstructed with manufacturer's iterative algorithm including attenuation correction, resolution recovery and eventually scatter correction, but without any regularization technique. Cold rod residual scatter fraction, hot and cold rod contrast recovery coefficient, coefficient of variation in phantom uniform areas and quantification accuracy using calibration with one of the homogeneous phantoms were, among others, computed.
Narrow symmetric photopeak-centred windows or fully asymmetric (≥ 140 keV) window led, on one hand, to decreased scatter residual fraction and sensitivity and, on the other hand, to increased noise, cold and hot recovery coefficients when compared to a standard 15-20% wide symmetric window. With a 6-7% wide symmetric window we obtained very comparable results to the dual energy window scatter correction used by the manufacturer for all measured parameters, but larger recovery coefficients especially for small hot objects in a cold background. Similar results were obtained with the fully asymmetric window at the cost of a higher noise level resulting from a drastic reduction of the sensitivity.
Narrow symmetric or asymmetric energy windows were found an interesting alternative to the standard dual energy window method to reject
Tc scattered photons. As a key feature, they allowed to avoid the erasing of small hot objects in a null background that was observed with the standard dual energy window scatter correction.