Performance Evaluation of the Inveon Dedicated PET Preclinical Tomograph Based on the NEMA NU-4 Standards
by
Chen, Mu
,
Stout, David B
,
Bao, Qinan
in
Animals
,
Diagnostic Equipment
,
Fluorodeoxyglucose F18
2009
The Inveon dedicated PET (DPET) scanner is the latest generation of preclinical PET systems devoted to high-resolution and high-sensitivity murine model imaging. In this study, we report on its performance based on the National Electrical Manufacturers Association (NEMA) NU-4 standards.
The Inveon DPET consists of 64 lutetium oxyorthosilicate block detectors arranged in 4 contiguous rings, with a 16.1-cm ring diameter and a 12.7-cm axial length. Each detector block consists of a 20 x 20 lutetium oxyorthosilicate crystal array of 1.51 x 1.51 x 10.0 mm elements. The scintillation light is transmitted to position-sensitive photomultiplier tubes via optical light guides. Energy resolution, spatial resolution, sensitivity, scatter fraction, and counting-rate performance were evaluated. The NEMA NU-4 image-quality phantom and a healthy mouse injected with (18)F-FDG and (18)F(-) were scanned to evaluate the imaging capability of the Inveon DPET.
The energy resolution at 511 keV was 14.6% on average for the entire system. In-plane radial and tangential resolutions reconstructed with Fourier rebinning and filtered backprojection algorithms were below 1.8-mm full width at half maximum (FWHM) at the center of the field of view. The radial and tangential resolution remained under 2.0 mm, and the axial resolution remained under 2.5-mm FWHM within the central 4-cm diameter of the field of view. The absolute sensitivity of the system was 9.3% for an energy window of 250-625 keV and a timing window of 3.432 ns. At a 350- to 625-keV energy window and a 3.432-ns timing window, the peak noise equivalent counting rate was 1,670 kcps at 130 MBq for the mouse-sized phantom and 590 kcps at 110 MBq for the rat-sized phantom. The scatter fractions at the same acquisition settings were 7.8% and 17.2% for the mouse- and rat-sized phantoms, respectively. The mouse image-quality phantom results demonstrate that for typical mouse acquisitions, the image quality correlates well with the measured performance parameters in terms of image uniformity, recovery coefficients, attenuation, and scatter corrections.
The Inveon system, compared with previous generations of preclinical PET systems from the same manufacturer, shows significantly improved energy resolution, sensitivity, axial coverage, and counting-rate capabilities. The performance of the Inveon is suitable for successful murine model imaging experiments.
Journal Article
Micro-chemical synthesis of molecular probes on an electronic microfluidic device
by
Sadeghi, Saman
,
van Dam, R. Michael
,
Dooraghi, Alex
in
Animals
,
Biological Sciences
,
Chemical reactions
2012
We have developed an all-electronic digital microfluidic device for microscale chemical synthesis in organic solvents, operated by electrowetting-on-dielectric (EWOD). As an example of the principles, we demonstrate the multistep synthesis of [18F]FDG, the most common radiotracer for positron emission tomography (PET), with high and reliable radio-fluorination efficiency of [18F]FTAG (88 ± 7%, n = 11) and quantitative hydrolysis to [18F]FDG (> 95%, n = 11). We furthermore show that batches of purified [18F]FDG can successfully be used for PET imaging in mice and that they pass typical quality control requirements for human use (including radiochemical purity, residual solvents, Kryptofix, chemical purity, and pH). We report statistical repeatability of the radiosynthesis rather than best-case results, demonstrating the robustness of the EWOD microfluidic platform. Exhibiting high compatibility with organic solvents and the ability to carry out sophisticated actuation and sensing of reaction droplets, EWOD is a unique platform for performing diverse microscale chemical syntheses in small volumes, including multistep processes with intermediate solvent-exchange steps.
Journal Article
Optical Characterization of AlAsSb Digital Alloy and Random Alloy on GaSb
by
Liang, Baolai
,
Chatziioannou, Arion
,
Huffaker, Diana
in
Alloying
,
Alloys
,
aluminum arsenide antimonide
2017
III-(As, Sb) alloys are building blocks for various advanced optoelectronic devices, but the growth of their ternary or quaternary materials are commonly limited by spontaneous formation of clusters and phase separations during alloying. Recently, digital alloy growth by molecular beam epitaxy has been widely adopted in preference to conventional random alloy growth because of the extra degree of control offered by the ordered alloying. In this article, we provide a comparative study of the optical characteristics of AlAsSb alloys grown lattice-matched to GaSb using both techniques. The sample grown by digital alloy technique showed stronger photoluminescence intensity, narrower peak linewidth, and larger carrier activation energy than the random alloy technique, indicating an improved optical quality with lower density of non-radiative recombination centers. In addition, a relatively long carrier lifetime was observed from the digital alloy sample, consistent with the results obtained from the photoluminescence study.
Journal Article
NEMA NU 4-2008 Comparison of Preclinical PET Imaging Systems
by
Goertzen, Andrew L.
,
Lecomte, Roger
,
Sossi, Vesna
in
Animals
,
Design
,
Image Processing, Computer-Assisted
2012
The National Electrical Manufacturers Association (NEMA) standard NU 4-2008 for performance measurements of small-animal tomographs was recently published. Before this standard, there were no standard testing procedures for preclinical PET systems, and manufacturers could not provide clear specifications similar to those available for clinical systems under NEMA NU 2-1994 and 2-2001. Consequently, performance evaluation papers used methods that were modified ad hoc from the clinical PET NEMA standard, thus making comparisons between systems difficult.
We acquired NEMA NU 4-2008 performance data for a collection of commercial animal PET systems manufactured since 2000: microPET P4, microPET R4, microPET Focus 120, microPET Focus 220, Inveon, ClearPET, Mosaic HP, Argus (formerly eXplore Vista), VrPET, LabPET 8, and LabPET 12. The data included spatial resolution, counting-rate performance, scatter fraction, sensitivity, and image quality and were acquired using settings for routine PET.
The data showed a steady improvement in system performance for newer systems as compared with first-generation systems, with notable improvements in spatial resolution and sensitivity.
Variation in system design makes direct comparisons between systems from different vendors difficult. When considering the results from NEMA testing, one must also consider the suitability of the PET system for the specific imaging task at hand.
Journal Article
Molecular imaging of small animals with dedicated PET tomographs
2002
Biological discovery has moved at an accelerated pace in recent years, with a considerable focus on the transition from in vitro to in vivo models. As a result, there has been a significant increase in the need to adapt clinical imaging methods, as well as for novel imaging technologies for biological research. Positron emission tomography (PET) is a clinical imaging modality that permits the use of positron-labeled molecular imaging probes for non-invasive assays of biochemical processes. The imaging procedure can be repeatedly performed before and after interventions, thereby allowing each animal to be used as its own control. Positron-labeled compounds that target a range of molecular targets have been and continue to be synthesized, with examples of biological processes ranging from receptors and synthesis of transmitters in cell communication, to metabolic processes and gene expression. In animal research, PET has been used extensively in the past for studies of non-human primates and other larger animals. New detector technology has improved spatial resolution, and has made possible PET scanning for the study of the most important modern molecular biology model, the laboratory mouse. This paper presents the challenges facing PET technology as applied to small animal imaging, provides a historical overview of the development of small animal PET systems, and discusses the current state of the art in small animal PET technology.
Journal Article
Quantitative Assessment of Longitudinal Metabolic Changes In Vivo after Traumatic Brain Injury in the Adult Rat using FDG-MicroPET
by
Osteen, Cheri L.
,
Cherry, Simon R.
,
Hovda, David A.
in
Animals
,
Autoradiography
,
Behavior, Animal
2000
With the advent and emerging importance of neurobiology and its relation to behavior, scientists desire the capability to apply noninvasive, quantitative imaging of neuronal activity to small rodents. To this end, the authors' laboratory has developed microPET, a high-resolution positron emission tomography (PET) scanner that is capable of performing in vivo molecular imaging at a resolution sufficient to resolve major structures in the rat brain. The authors report in this article that, in conjunction with 2-[18F]fluoro-2-deoxyglucose (FDG), microPET provides accurate rates of cerebral glucose metabolism (59.7 to 108.5 μmol/100 g·min) in conscious adult rats as validated by within-subject autoradiographic measurements (59.5 to 136.2 μmol/100 g·min;r = 0.88; F[1,46] = 168.0;P < 0.001). By conducting repeated quantitative scanning, the authors demonstrate the sensitivity and accuracy of FDG-microPET to detect within-subject metabolic changes induced by traumatic brain injury. In addition, the authors report that longitudinal recovery from traumatic brain injury-induced metabolic depression, as measured by quantitative FDG-microPET, is significantly correlated (r = 0.65;P < 0.05) to recovery of behavioral dysfunction, as assessed by the Morris Water Maze performance of the same rats, after injury. This is the first study to demonstrate that FDG-microPET is quantitative, reproducible, and sensitive to metabolic changes, introducing a new approach to the longitudinal study of small animal models in neuroscience research.
Journal Article
Multimodality Imaging of Tumor Xenografts and Metastases in Mice with Combined Small-Animal PET, Small-Animal CT, and Bioluminescence Imaging
2007
Recent developments have established molecular imaging of mouse models with small-animal PET and bioluminescence imaging (BLI) as an important tool in cancer research. One of the disadvantages of these imaging modalities is the lack of anatomic information. We combined small-animal PET and BLI technology with small-animal CT to obtain fusion images with both molecular and anatomic information.
We used small-animal PET/CT and BLI to detect xenografts of different cell lines and metastases of a melanoma cell line (A375M-3F) that had been transduced with a lentiviral vector containing a trimodality imaging reporter gene encoding a fusion protein with Renilla luciferase, monomeric red fluorescent protein, and a mutant herpes simplex virus type 1 thymidine kinase.
Validation studies in mouse xenograft models showed a good coregistration of images from both PET and CT. Melanoma metastases were detected by 18F-FDG PET, 9-[4-(18)F-fluoro-3-(hydroxymethyl)butyl]guanine (18F-FHBG) PET, CT, and BLI and confirmed by ex vivo assays of Renilla luciferase and mutant thymidine kinase expression. 18F-FHBG PET/CT allowed detection and localization of lesions that were not seen on CT because of poor contrast resolution and were not seen on 18F-FDG PET because of higher background uptake relative to 18F-FHBG.
The combination of 18F-FHBG PET, small-animal CT, and BLI allows a sensitive and improved quantification of tumor burden in mice. This technique is potentially useful for the study of the biologic determinants of metastasis and for the evaluation of novel cancer treatments.
Journal Article
A Deformable Atlas of the Laboratory Mouse
2015
Purpose
This paper presents a deformable mouse atlas of the laboratory mouse anatomy. This atlas is fully articulated and can be positioned into arbitrary body poses. The atlas can also adapt body weight by changing body length and fat amount.
Procedures
A training set of 103 micro-CT images was used to construct the atlas. A cage-based deformation method was applied to realize the articulated pose change. The weight-related body deformation was learned from the training set using a linear regression method. A conditional Gaussian model and thin-plate spline mapping were used to deform the internal organs following the changes of pose and weight.
Results
The atlas was deformed into different body poses and weights, and the deformation results were more realistic compared to the results achieved with other mouse atlases. The organ weights of this atlas matched well with the measurements of real mouse organ weights. This atlas can also be converted into voxelized images with labeled organs, pseudo CT images and tetrahedral mesh for phantom studies.
Conclusions
With the unique ability of articulated pose and weight changes, the deformable laboratory mouse atlas can become a valuable tool for preclinical image analysis.
Journal Article
Adenovirus-mediated gene expression imaging to directly detect sentinel lymph node metastasis of prostate cancer
by
Burton, Jeremy B
,
Mulholland, David J
,
Koh, Sok Boon S
in
Adenoviridae - genetics
,
Adenoviridae - metabolism
,
Adenovirus
2008
The degree of lymph-node metastasis in prostate cancer is crucial for both staging the disease and planning treatment. Here, Burton and colleagues describe a one-step, non-invasive imaging technology using prostate-specific adenoviral vectors that express imaging reporter genes. This set-up specifically and accurately detects lymph-node metastases in a model of human prostate cancer and eliminates the need for invasive lymphadenectomy required by the current lymphoscintigraphy method.
The accurate assessment of nodal involvement in prostate cancer is crucial to planning treatment, yet there is a shortage of noninvasive imaging techniques capable of visualizing nodal lesions directly. This study demonstrates the feasibility of using recombinant human adenoviral vectors to detect nodal metastases in a human prostate cancer model. This was achieved by the prostate-restricted expression of optical and positron emission tomography (PET) imaging reporter genes by the viral vector coupled with the innate lymphotropic properties of adenovirus. We show that peritumoral administration of these vectors results in the direct detection of reporter gene expression in metastatic lesions within sentinel lymph nodes. Notably, this approach parallels the current lymphoscintigraphy method but enables the direct PET visualization of sentinel lymph node metastases, eliminating the need for invasive lymphadenectomy. These findings may lead to more effective diagnostic and therapeutic strategies for individuals with advanced-stage prostate cancer.
Journal Article
Mouse Atlas Registration with Non-tomographic Imaging Modalities—a Pilot Study Based on Simulation
by
Wang, Hongkai
,
Stout, David B.
,
Chatziioannou, Arion F.
in
Anatomy, Artistic
,
Animals
,
Atlases as Topic
2012
Purpose
This study investigates methodologies for the estimation of small animal anatomy from non-tomographic modalities, such as planar X-ray projections, optical cameras, and surface scanners. The key goal is to register a digital mouse atlas to a combination of non-tomographic modalities, in order to provide organ-level anatomical references of small animals in 3D.
Procedures
A 2D/3D registration method was developed to register the 3D atlas to the combination of non-tomographic imaging modalities. Eleven combinations of three non-tomographic imaging modalities were simulated, and the registration accuracy of each combination was evaluated.
Results
Comparing the 11 combinations, the top-view X-ray projection combined with the side-view optical camera yielded the best overall registration accuracy of all organs. The use of a surface scanner improved the registration accuracy of skin, spleen, and kidneys.
Conclusions
The methodologies and evaluation presented in this study should provide helpful information for designing preclinical atlas-based anatomical data acquisition systems.
Journal Article