False positive results of Bowie and Dick type test used for hospital steam sterilizer with slower come-up ramps: A case study
by
Bronzatti, Jeane Aparecida Gonzalez
, Bruna, Camila Quartim de Moraes
, Laranjeira, Paulo Roberto
, de Souza, Rafael Queiroz
, Lusignan, Viktoriya
, Graziano, Kazuko Uchikawa
in
Biology and Life Sciences
/ Case reports
/ Case studies
/ Comparative analysis
/ Configurations
/ Criteria
/ Engineering and Technology
/ Equipment and Supplies, Hospital - microbiology
/ Equipment Contamination - prevention & control
/ Gases
/ Health risks
/ Humans
/ Medical devices
/ Medical electronics
/ Medical equipment
/ Medical equipment industry
/ Medicine and Health Sciences
/ Methods
/ Nursing schools
/ People and Places
/ Physical Sciences
/ Quality Control
/ Research and Analysis Methods
/ Science Policy
/ Steam
/ Sterilization
/ Sterilization - instrumentation
/ Sterilization - methods
/ Temperature
2020
Hey, we have placed the reservation for you!
By the way, why not check out events that you can attend while you pick your title.
You are currently in the queue to collect this book. You will be notified once it is your turn to collect the book.
Oops! Something went wrong.
Looks like we were not able to place the reservation. Kindly try again later.
Are you sure you want to remove the book from the shelf?
False positive results of Bowie and Dick type test used for hospital steam sterilizer with slower come-up ramps: A case study
by
Bronzatti, Jeane Aparecida Gonzalez
, Bruna, Camila Quartim de Moraes
, Laranjeira, Paulo Roberto
, de Souza, Rafael Queiroz
, Lusignan, Viktoriya
, Graziano, Kazuko Uchikawa
in
Biology and Life Sciences
/ Case reports
/ Case studies
/ Comparative analysis
/ Configurations
/ Criteria
/ Engineering and Technology
/ Equipment and Supplies, Hospital - microbiology
/ Equipment Contamination - prevention & control
/ Gases
/ Health risks
/ Humans
/ Medical devices
/ Medical electronics
/ Medical equipment
/ Medical equipment industry
/ Medicine and Health Sciences
/ Methods
/ Nursing schools
/ People and Places
/ Physical Sciences
/ Quality Control
/ Research and Analysis Methods
/ Science Policy
/ Steam
/ Sterilization
/ Sterilization - instrumentation
/ Sterilization - methods
/ Temperature
2020
Oops! Something went wrong.
While trying to remove the title from your shelf something went wrong :( Kindly try again later!
Do you wish to request the book?
False positive results of Bowie and Dick type test used for hospital steam sterilizer with slower come-up ramps: A case study
by
Bronzatti, Jeane Aparecida Gonzalez
, Bruna, Camila Quartim de Moraes
, Laranjeira, Paulo Roberto
, de Souza, Rafael Queiroz
, Lusignan, Viktoriya
, Graziano, Kazuko Uchikawa
in
Biology and Life Sciences
/ Case reports
/ Case studies
/ Comparative analysis
/ Configurations
/ Criteria
/ Engineering and Technology
/ Equipment and Supplies, Hospital - microbiology
/ Equipment Contamination - prevention & control
/ Gases
/ Health risks
/ Humans
/ Medical devices
/ Medical electronics
/ Medical equipment
/ Medical equipment industry
/ Medicine and Health Sciences
/ Methods
/ Nursing schools
/ People and Places
/ Physical Sciences
/ Quality Control
/ Research and Analysis Methods
/ Science Policy
/ Steam
/ Sterilization
/ Sterilization - instrumentation
/ Sterilization - methods
/ Temperature
2020
Please be aware that the book you have requested cannot be checked out. If you would like to checkout this book, you can reserve another copy
We have requested the book for you!
Your request is successful and it will be processed during the Library working hours. Please check the status of your request in My Requests.
Oops! Something went wrong.
Looks like we were not able to place your request. Kindly try again later.
False positive results of Bowie and Dick type test used for hospital steam sterilizer with slower come-up ramps: A case study
Journal Article
False positive results of Bowie and Dick type test used for hospital steam sterilizer with slower come-up ramps: A case study
2020
Request Book From Autostore
and Choose the Collection Method
Overview
To determine if the standardized B&D type test for hospital steam sterilizer is correctly indicating cycle failures in slower come-up ramps cycles.
Two commercially available B&D type test were challenged in a quality control sterilizer. A common failure was simulated in triplicate cycles, using a standardized cycle configuration, and then compared to triplicate cycles of a common cycle configuration. Tests procedures were conducted according to B&D manufacturer test standard and results were compared to standardized endpoint specifications.
We found that B&D type tests are only capable of detecting the presence of non-condensable gases if the sterilization equipment is adjusted to meet all the cycle requirements of the test. False positive results were obtained with come-up ramp time of 3 min. Correct results were only obtained with come-up ramp of 1.7-1.9 min.
Until the ISO 17665 and AAMI ST-79 standards are revised, equipment qualification experts should observe come-up time duration criterion for B&D type test cycles according to ISO 11140-4. Sterile Processing Department professionals must add the come-up ramp criterion to cycle evaluation before clearing the equipment for routine use. This will allow B&D correct performance, reducing the infection risk from unsterilized medical device.